SBRT/SRS patient-specific QA using GAFchromic™ EBT3 and FilmQA™ Pro software
Giuseppe Stella, Nina Cavalli, Elisa Bonanno, Lucia Zirone, Giuseppina Rita Borzì, Martina Pace, Andrea Girlando, Anna.M. Gueli and Carmelo Marino
The aim of this work is to verify the potential use of GAFchromic™ EBT3 and FILMQA™ pro software for patient specific quality assurance (QA) for stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT) treatment plans in clinical routine use. In particular, encephalic, pulmonary and lymph node treatments plans were selected for this study. The agreement between the calculated and measured dose distributions were evaluated in terms of ɣ index with 3%3mm, 2%2mm, 1.5%1.5mm and 3%1.5mm criteria. The obtained results were then compared to the routine pre-treatment verification method which uses electronic portal imaging device (EPID) and EPIQA analysis software. EBT3-FilmQA method results show a mean ɣ index passing rate >95% with 2%1.5mm analysis criteria and an improvement of about 7% compared with EPID-EPIQA method results.
Keywords: 2D dosimetry, patient dose, treatment planning system, small fields
After payment has been processed for your order of a digital copy (PDF) of this article, you will see a download link on your completed order page and also receive an email containing a download link. The links, which will enable you to download one copy of the article, will expire after 24 hours.