Additional Information

Site Information

 Loading... Please wait...
  • Image 1

JRSBRT 5.2, p. 123-129

$40.00
Quantity:

Product Description

Targeting accuracy at couch kick for a frameless image guided radiosurgery system
Yimei Huang, Bo Zhao, Joshua Kim, Ning Wen, Indrin J. Chetty and Salim Siddiqui

Purpose/Objectives: Targeting accuracy at all possible couch angles needs to be carefully evaluated prior to initiating a frameless image-guided stereotactic radiosurgery program on a Linac for treating functional disorders such as trigeminal neuralgia. In this study, we report positioning accuracy with stereoscopic x-ray imaging over the complete range of couch rotation using anthropomorphic head phantoms.

Materials/Methods: An anthropomorphic head phantom with three 5 mm tungsten BBs as hidden targets was CT simulated. A group of 7 arcs was planned with couch angles from 0 to 90° in 15° increments. A pair of stereoscopic x-ray images that auto-matches to planning CT according to bony anatomy was utilized to position each BB to machine isocenter at all planned couch angles. Targeting accuracy was measured by stereoscopic x-ray imaging of the BB itself, which provides the distance from the centroid of BB to the x-ray imaging isocenter. For each BB, the hidden target test was repeated 5 times at couch 0° and 3 times at other couch angles, resulting in a total of 69 measurements, each with random initial setup deviation. Following the same workflow, a second anthropomorphic head phantom with two 5 mm BBs was utilized to evaluate localization accuracy at couch angles of 0 through 270° in 15° increments, resulting in another 18 measurements.

Results: Residual setup deviation following image guidance in the first head phantom was 0.6±0.1, 0.4±0.1, and 0.4±0.1 mm, respectively, at the three BBs, and 0.5±0.1 mm overall (N=69). Comparable results of 0.4±0.1 mm (N=18) were achieved with the second head phantom. Combining results from both phantoms, the targeting accuracy was 0.5±0.1 mm (range 0.2-0.8 mm). No apparent correlation was observed between targeting accuracy and couch rotation.

Conclusions: Accurate positioning within 1 mm can be achieved with stereoscopic x-ray imaging at any couch angle. Frameless image-guided stereotactic radiosurgery could achieve targeting accuracy similar to that of frame-based systems for high dose treatment of trigeminal neuralgia.

Keywords: frameless radiosurgery, trigeminal neuralgia, image-guided radiosurgery

After payment has been processed for your order of a digital copy (PDF) of this article, you will see a download link on your completed order page and also receive an email containing a download link. The links, which will enable you to download one copy of the article, will expire after 24 hours.

Find Similar Products by Category