Additional Information

Site Information

 Loading... Please wait...
  • Image 1

JRSBRT 6.4, p. 295-301

$40.00
Quantity:

Product Description

Accuracy and efficiency of image-guided radiation therapy (IGRT) for preoperative partial breast radiosurgery
Sua Yoo, Jennifer O’Daniel, Rachel Blitzblau, Fang-Fang Yin and Janet K. Horton

Objective: To analyze and evaluate accuracy and efficiency of IGRT process for preoperative partial breast radiosurgery.

Methods: Patients were initially setup with skin marks and 5 steps were performed: (1) Initial orthogonal 2D kV images, (2) pre-treatment 3D CBCT images, (3) verification orthogonal 2D kV images, (4) treatment including mid-treatment 2D kV images (for the final 15 patients only), and (5) post-treatment orthogonal 2D kV or 3D CBCT images. Patient position was corrected at each step to align the biopsy clip and to verify surrounding soft tissue positioning.

Results: The mean combined vector magnitude shifts and standard deviations at the 5 imaging steps were (1) 0.96 ± 0.69, (2) 0.33 ± 0.40, (3) 0.05 ± 0.12, (4) 0.15 ± 0.17, and (5) 0.27 ± 0.24 in cm. The mean total IGRT time was 40.2 ± 13.2 minutes. Each step was shortened by 2 to 5 minutes with improvements implemented. Overall, improvements in the IGRT process reduced the mean total IGRT time by approximately 20 minutes. Clip visibility was improved by implementing oblique orthogonal images.

Conclusion: Multiple imaging steps confirmed accurate patient positioning. Appropriate planning and imaging strategies improved the effectiveness and efficiency of the IGRT process for preoperative partial breast radiosurgery.

Keywords: breast radiosurgery, IGRT, accuracy, efficiency

After payment has been processed for your order of a digital copy (PDF) of this article, you will see a download link on your completed order page and also receive an email containing a download link. The links, which will enable you to download one copy of the article, will expire after 24 hours.

Find Similar Products by Category